Customization: | Available |
---|---|
Customized: | Non-Customized |
Certification: | CE, ISO13485 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Multi-functional Portable DR Mobile X-ray Machine FN-MTM200
Features:
1. The portable Digital X-ray system is small and lightweight, with four universal wheels which can achieve real freedom of movement.
2. Tube head has LCD touch screen, which can select body position and focus, and adjust kV, mAs, mA and ms. This smart design is convenient for positioning and adjusting exposure parameters, which greatly improves the work efficiency ofradiographers.
3. There is a large storage space for the flat panel detector. A password lock is designed to help the FPD management.
4. Free stop tube arm for fast and accurate patient positioning.
Configurations | ||||||
No. | Brand | Model | Product Name | Quantity | ||
1 | Foinoe | TCQ -IV | Flat Panel Detector | 1 | ||
Foinoe | TCQ -III | Flat Panel Detector | Optional | |||
2 | Foinoe | DXRayDiagnost | DXRay Diagnostic Workstation | 1 | ||
3 | Foinoe | FSQ -5 -PD5E | High Frequency Generator | 1 | ||
4 | Foinoe | KL10SB | X-Ray Tube | 1 | ||
5 | Foinoe | Collimator | 1 | |||
6 | Foinoe | MTM200 | Mobile Rack | 1 | ||
7 | Dell | Ins 15 -3501 -R1605s | Laptop | 1 | ||
SYSTEM CONFIGURATION | |
WIFI FPD Detector | |
Angell DXRayDiagnost | |
5.6kW High frequency generator | |
X-Ray tube | |
Mobile rack | |
Static WIFI Flat Panel Detector | |
Detector: Wireless detector Detector technology: Cesium iodide (CsI)scintillator coupled to TFT matrix with amorphous silicon technology Image Size: 35×43cm (14×17 inch) Pixel Matrix: 2560×3072 Pixel size: 140μm A/D Conversion: 16 bits DQE: 2μGy at 0 lp/mm 65% Spatial Resolution: 3.6 LP/mm Image Acquisition Time: 2s Image Processing Time: 3s Recommended cycle time:10s Communication: Ethernet Adapter Input: AC 100-240V,50-60Hz Adapter Output: DC 24V, 60W Dimensions: 38×46×1.5 cm |
Weight: 3.3 kg Max. load capacity 135 kg with patient lying on it; 100 kg with patient standing Battery: Lithium ion, build in, rechargeable, 3000mAh 2 pieces Charging time: 240 minutes for 100 % battery capacity Battery Standby Time: 12H Housing Material: Carbon, Alloy Water Tightness: IP54 Operating Environment: 5-35 ºC,10-75% RH |
DXRay Diagnostic Workstation |
Dell laptop CPU: Intel i5 2.4GHz Memory: 16GB ,Hard Drive: 512G SSD Display: 15.6inch; Resolution: 1920*1080 pixels DXRayDiagnost software Case management Case management function: including the management of patient information, exam information and images. Worklist enquiry service following standard DICOM3.0 may inquiry from HIS/PACS and download case information. Image acquisition Acquisition system: positive and negative film image capture can be configured. Real time automatic window width level adjustment. Real time ROI (region of interests) clipping. Real time enhancement. Selecting image and rotation according to different body postures. Displaying patient information, examination information, device information and image information Image processing Adjustment of window width, window level, Gamma, and LUT curve. Positive and negative film switchover, image zoom, translation, mirror image, rotation, magnified display, flip, invert. Image enhancement, noise reduction: Scheme and level of image enhancement and noise reduction can be adjusted according to different body postures. Image labeling function: including drawing straight line, rectangle, polygon, arrow and text. Real time reminding the free space of image disk ofthe system. RT prompt system can be used for store size ofthe disk space of image. Image output |
Sending SCU with DICOM: support sending image to any PACS and workstations following Standard DICOM3.0 through internet of 5G speed WIFI or Lan. Receiving SCP server with DICOM: support receiving image following Standard DICOM3.0 from any workstation. DICOM images can be saved as DICOM printer, folder, JPG, BMP, CD/DVD through the driver. Emergency mode: it can supply emergency exposure without registration Check-up mode: it can import and register patient information by excel Reject analysis: Image cancel / delete image / Statistic ofuse Reject analysis: Image cancel / delete image / Statistic of u |
||
5.6KW High Frequency Generator | ||
Nominal supply voltage: 100~264VAC Working Frequency: 80kHz~200kHz Maximum output power: 5.6KW Nominal Power: 5.6KW Tube Voltage Setting Range and Digital Display: 40 to 125kV, in 1kV increments Tube current: 5 to 100mA Current-Time Product Setting Range and Digital Display: 0.1-320mAs Exposure time range: 1ms- 10000ms Size: 383mm*240mm*214mm Weight: 14.5kg±0.5kg |
||
X-Ray tube (KL10SB) | ||
Target angle: 15° Focal Spot Sizes: 0.6 / 1.8 mm Max. Anode Heat Content: 20kJ Max. heat content: 40 kJ |
||
Collimator | ||
Max. Irradiation field (SID=100cm): rectangular irradiation field 430mm×430mm Power supply: 24VAC 6A Average illumination (at SID=100cm): 160 lx or more |
||
Mobile Rack | ||
Max. distance from the tube focus to the ground (SID): 2000mm Tube rotation around support arm: 360° |
Tube rotation around tube axis: 350° Tube head LCD Screen console could select Focus, and adjust kV, mA, mAs, and ms; also indicates exposure status and displays Error code. System size (L*W*H): 795mm*550mm*1450mm Total weight: 67kg (for whole machine) |