Digital Medical X-ray Radiographic System Multi-functional Diagnostic Dynamic DR Foinoe FN-650MA-2 Remote Table Series
FN-650MA-2 PACKING LIST | ||||||
Systems | Parts | Quantity | Model | |||
Detector | Dynamic Flat Panel Detector | 1 | / | |||
DXRay Diagnostic Workstation | 1 | / | ||||
X-Ray System | High Frequency Generator | 1 | / | |||
Control Cabinet | 1 | / | ||||
X-Ray Tube | 1 | / | ||||
High Voltage Cables | 2 | / | ||||
Collimator | 1 | / | ||||
Motion Control System | Remote Control Table | 1 | / | |||
Remote-Control | 1 | / | ||||
Mobile Remote Table(Optional) | 1 | / | ||||
Accessories | Grid | 1 | / | |||
Report Laser Printer | 1 | / | ||||
Microphone/Speaker | 1 | / |
FN-650MA-2 SPECIFICATIONS | |||||
Part | Remark | ||||
Dynamic Flat Panel Detector | Dynamic Flat Panel Detector Type | a-Si Flat Panel Detector | |||
Scintillator Screen Type | Csl (Cesium lodide) | ||||
Effective View Field | 17 × 17 in. (43 × 43 cm) | ||||
Static Pixel Matrix | 3072 x 3072 | ||||
Dynamic Pixel Matrix | 1536 x 1536 | ||||
Spot-Film Pixel Matrix | 3072 x 3072 | ||||
Spot-Film Preparation Time | ≤0.8s | ||||
Output Grayscale | 16bit | ||||
Preview Image Time | ≤3s | ||||
Maximum Spatial Resolution | 3.5 lp/mm | ||||
Detector Movement Range | ≥950mm | ||||
X-ray Tube | Tube Focus | 0.6mm/1.0mm | |||
Max Output Voltage | 150kV | ||||
Heat Capacity | 330kHU | ||||
Output Power | 30kW/50kW | ||||
High Frequency Generator | Power Requirement | 380V with 3 Phases Wires | |||
Maximum Output Power | 50kW | ||||
Output Voltage for Radiography | 40kV-150kV | ||||
Output Voltage for Fluoroscopy | 40kV-120kV | ||||
Output Current for Radiography | 10mA-650mA | ||||
Output Current for Fluoroscopy | 0.5mA-10mA | ||||
Range of mAs | 0.4mAs-630mAs | ||||
Input Power Frequency | 50Hz/60Hz | ||||
Auto Brightness Selection Function(ABS) | |||||
Remote Control Table | Table Size | 2100mm x 880mm x 730mm | |||
Table Lateral Movement | 250mm | ||||
SID | 1000-1800mm | ||||
Table Rotate Range | -15° - +90° | ||||
Tube Column Rotate Range | -35° - +35° | ||||
High Voltage Cables of X-ray Machine | Length: 8m | ||||
Rotation Angle of Foot Pedal | 360° | ||||
One-click in Place Function | Chest Position or Table Position | ||||
Remote-Control | 1.One-click in Place Function:One-click to Chest Position or Table Position | ||||
2.Exposure fluoroscopy switch: Exposure and fluoroscopy can be prohibited | |||||
to ensure the safety of doctors and prevent accidental exposure | |||||
Mobile Remote Table (Optional) | Height range | 770mm - 1170mm | |||
1.One-click in Place Function:One-click to Chest Position or Table Position | |||||
2.Exposure fluoroscopy switch: Exposure and fluoroscopy can be prohibited to ensure the safety of doctors and prevent accidental exposure |
|||||
Grid | Grid Ratio | 10:01 | |||
Grid Focus | 120 cm | ||||
Support manual removal | |||||
Collimator | Power | 150W(24VAC) | |||
Auto Collimation | Yes | ||||
Inherent Filtering | ≤1.5mmAl@120KV (XS-D2) | ||||
Workstation (can be customized) | CPU | INTEL-I5 | |||
RAM | 4G | ||||
Hard Disk | 500G | ||||
Display | 1280*1024 pixels or above | ||||
CD/DVD recording/burning | |||||
Software | 1.Patient management: Manual registration, Automatic Query from Worklist | ||||
2.Image acquisition: Automatic Window Adjustment, Automatic Sending,Static Image Acquisition, Dynamic Image Acquisition,Video Save, Playback,Bone Stitching(Optional) |
|||||
3.Image processing: Image Correction, Image Clipping,Automatic Image Segmentation,EAE Image Enhancement Processing , IEQ Image Processing |
|||||
4.Image viewer: Window/Level Adjustment, Image Flip, Image Rotation, Image Magnification, Restore |
|||||
5.Medical report: Automatic Loading of Patient Information, Expert Template | |||||
6.Film printing: Supports Standard DICOM 3.0 Printer | |||||
7.DICOM transmission: Complies with the standard DICOM 3.0 PACS and Workstation |
|||||
8. Fluoroscopy amplification: 3 step : 17/15,17/12,17/9. | |||||
9. Anatomically programmed radiography (APR) : auto pre-set parameter link to the chosen body part |
|||||
10. Automatical collimator: After choose the body part, collimator can adjust the suitable window size, changing SID could aslo bring automatical adjustment of collimator |